10 - 47
HAPSITE ER Operating Manual
The Elimination Rules section gives parameters for peaks to be reported. There
are three options. See Figure 10-52.
Figure 10-52 Elimination Rules Window
Keep no more than. . . . . . . . . . . . . This check box determines the number of
peaks that will be displayed. If ten is
selected, the ten peaks with the largest areas
will be displayed.
Keep peaks above . . . . . . . . . . . . . When this box is checked, only peaks above
the desired percent intensity will be
displayed. Fifteen is generally used as the
low end for intensity.
Maximum NIST Hits . . . . . . . . . . . . This value is the number of matches that will
be reported by NIST.
NOTE: This section has a Reset button which will reset the entered values to the
default settings.
10.10.4 Alarm
On the main Method Editor Search page, the Alarm option in the Library
Actions box will activate when a calibrated library has been saved to the method.
To enable the Alarm option
1 Check the Alarm box. See Figure 10-53.