EasyManua.ls Logo

Inficon HAPSITE ER - Installing the Gas Canisters

Inficon HAPSITE ER
432 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3 - 7
HAPSITE ER Operating Manual
4 Insert the Valco connector into the top of HAPSITE ER. Screw the Valco
connector into place. (See Figure 3-6.)
Figure 3-6 Valco Connector
NOTE: Save all of the port caps for further use.These caps are necessary when
decontaminating HAPSITE ER. Spare caps are provided in the Ship Kit.
3.2.2 Installing the Gas Canisters
CAUTION
Do not open the front panel in a contaminated area.
The carrier and internal standard gas canisters must be installed inside
HAPSITE ER prior to sampling. See section 2.4.1, Gas Chromatograph, on page
2-3 for information on the gas canisters. Follow the instructions below to install the
gas canisters into the spring-loaded slots inside the unit.
1 Open the panel by placing thumbs on the top of panel and pulling downwards.
This technique avoids damaging the sealing gasket with fingernails. (See
Figure 3-7.)
Figure 3-7 Opening the Front Panel

Table of Contents

Related product manuals