Inficon HAPSITE ER - Testing Battery; Installing the Battery

Inficon HAPSITE ER
432 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3 - 19
HAPSITE ER Operating Manual
Red . . . . . . . . . . . . . . . . . . . . . . . . . The receptacle and/or the battery, if one is
installed is experiencing a problem. A
flashing red light indicates that the charger
cannot communicate with the battery.
Off . . . . . . . . . . . . . . . . . . . . . . . . . . The receptacle is ready to charge a battery. If
the indicator remains extinguished when a
battery is inserted, the battery is severely
depleted. In this case, leave the battery in the
receptacle and unplug the power cord.
Reconnect the power cord and the battery
will start to charge.
3.3.4.1 Testing Battery
1 To test a battery, press the TEST button on the end of the battery.
(See Figure 3-33.)
Figure 3-33 Battery TEST Button
2 In the elongated triangle, green lighted numbers will be displayed. The highest
illuminated number indicates the remaining percentage of battery charge,
which is reported in 20% increments.
NOTE: If OVER is illuminated, the battery is fully charged.
3.3.5 Installing the Battery
1 Insert a fully charged battery by sliding it into the rectangular opening to the left
of the gas canisters. The battery should be inserted with the release arrow
pointing towards the release button. (See Figure 3-34.)
Figure 3-34 Inserting Battery