9 - 14
HAPSITE ER Operating Manual
4 On the System Setup window, verify the Access Level is set to Advanced.
(See Figure 9-26.)
Figure 9-26 Verify access level
9.4.6 Manual Tune
1 Double click the icon. (See Figure 9-27.)
Figure 9-27 Tune icon
2 Select default.tun from the tuning pop-up window.
3 Click OK. It will take 10 to 20 seconds to initialize. During that time the
Emission and EM boxes turn green. (See Figure 9-28.)
NOTE: The NEG (non-evaporable getters) is consumed while Manual Tune is
open.
4 Adjust the tune parameters as discussed above until all values are within
range.