4 - 12
HAPSITE ER Operating Manual
4 HAPSITE ER will re-run the cleanout as part of its preparation.
5 If the failure box appears again, check the concentrator to verify that it is not
cracked or chipped. Also, try re-installing the concentrator to ensure that it is
properly seated.
4.1.4 Quick Reference SOP - Heat-up and Tune
CAUTION
Do not open the front panel in a wet or contaminated
area.
1 Insert the internal standard and carrier gas canisters.
2 Insert a charged battery.
3 Connect the AC to DC power converter power supply.
4 Verify that the appropriate sample configuration (i.e., concentrator) is installed.
5 Press the POWER button on the front panel.
6 HAPSITE ER will heat the necessary components and perform AutoTune. A
prompt to run SURVEY or ANALYZE will appear when HAPSITE ER is ready
to run a sample.
7 If the default method is not the desired method, touch STOP PREPARE.
8 Touch SELECT METHOD. Highlight the desired method. Touch Select.
NOTE: If connecting wirelessly to the laptop, see Chapter 5, Communications
and Touch Screen Options.
NOTE: When the SYSTEM READY message is displayed, touch either RUN
SURVEY or RUN ANALYZE. If using the push buttons, push SURVEY
RUN or ANALYZE RUN.