10 - 25
HAPSITE ER Operating Manual
10.8 Full Scan Page
The Full Scan page sets the mass ranges for the method. The set can be assigned
a Scan Set Name for easy identification, if desired. The Filament Delay, from the
Inlet Page (see section 10.6.3, Scan Events, on page 10-19), is also shown on the
Full Scan page. Changing the Filament Delay on this page may require changes
to the Inlet Page.
Figure 10-30 Method Editor Full Scan Page
The following Mass Spectrometer parameters can be programmed:
Start Mass . . . . . . . . . . . . . . . . . . . . The mass at which the Mass Spectrometer
will start to scan. The starting mass can be
set from 1 to 300 amu, however values
between 40-50 are recommended to avoid
detection of high-volatility interferences