EasyManua.ls Logo

Inficon HAPSITE ER - Set Access Level to Advanced

Inficon HAPSITE ER
432 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
9 - 13
HAPSITE ER Operating Manual
9.4.5 Set Access Level to Advanced
The Access Level must be set to Advanced to manually tune the instrument. The
Set Access Level... is used to change ER IQ/Plus IQ user mode.
Normal Mode: Allows access to default methods and data analysis
Advanced Mode: Allows access to method editing, manual tuning, file transfer,
and file deletion.
1 On the System Setup window Tools Menu, click Set Access Level....
(See Figure 9-24.)
Figure 9-24 Set Access Level
2 On the Change Access Level window, select Advanced in the Requested
Access Level box. (See Figure 9-25.)
NOTE: Access level is not password protected in the default settings.
Advanced mode may be password protected if desired.
Figure 9-25 Change Access Level window
3 Click OK.

Table of Contents

Related product manuals