7 - 13
HAPSITE ER Operating Manual
7.3.4 Tools Menu
The Tools menu is shown below. (See Figure 7-22.)
Figure 7-22 Tools menu
System Setup closes the System Setup screen.
System Properties functions identically to the System Properties option in the
System menu. Refer to section 7.3.3 on page 7-6.
Sensor Properties functions identically to the Properties option in the System
Menu. Refer to section 7.3.3 on page 7-6 for further information.
7.3.4.1 Set Access Level
In the Set Access Level option there are two user levels which can be set in
ER IQ, Normal and Advanced. Neither access level has a factory set password.
Normal level allows users to run samples, view results and perform basic
operations with HAPSITE ER.
Advanced allows access to all user operations. This includes all normal user
functions plus method creation and editing, file deletion, changing alarm
parameters and changing network settings.
To restrict access to advanced functions, an advanced user password can be set.
Once the password is set, it must be entered each time the ER IQ program is
opened, or whenever the access level is changed from normal to advanced. See
section 7.3.4.1.2 on page 7-15.