10.10 Search Page
The Search Page sets the necessary parameters to qualify and quantify data. To
quantify data, a calibration library must be created. See Figure 10-40. See Chapter
11, Calibration for instructions on creating a calibration library.
Figure 10-40 Method Editor Search Page
There are four choices in the Search Mode drop down menu.
No Search . . . . . . . . . . . . . . . . . . . . If this option is selected, a library search will
not be conducted and a report will not be
displayed on the front panel at the end of the
run.
NOTE: SIM Methods only allow No Search as the search option.
Qualitative. . . . . . . . . . . . . . . . . . . . Searches AMDIS during a run to provide
near real-time identifications. A report will be
generated at the end of the run and can be
viewed on the front panel display.