10 - 36
HAPSITE ER Operating Manual
8 It is recommended that the Dwell Time is adjusted until the Round Trip is
approximately one second. (See Figure 10-39.)
Figure 10-39 Adjusting Dwell Time
9 Repeat Step 4 through Step 8 if successive chemicals are desired.
NOTE: The End Time for the final SIM compound must be the same as the
end run time for the method.
NOTE: AMDIS and the SearchNIST/User libraries are not available when
using a SIM method.