9 - 9
HAPSITE ER Operating Manual
Status Column
The Status Column provides categorical view of the actual percentages. (See 
Figure 9-18.) The values OK, OK Low, OK High, Low or High indicate whether 
the actual percentage for that ion are within best range (OK) or outside (Low or 
High). Values of High or Low are not acceptable. Adjust the corresponding mass 
fragments.
Figure 9-18  Status column
9.4.3  Inputs
Target Resolution
Resolution refers to the width of the measured mass peak in amu. The range for 
the Target Resolution is 0.85 and 1.10. Adjusting Resolution influences the Actual 
Percentages. (See Figure 9-19.)
Figure 9-19  Target Resolution
Wider resolutions mean larger percentages (measuring over a wider section of the 
MS window). As these are adjusted, watch and observe the changes to the 
percentages.
% of Fragment 55 is above 
the appropriate range
% of Fragment 282 is within the 
appropriate range, no need for adjustment
% of Fragment 213 is outside the optimal 
range; adjustment is not necessary