8 - 42
HAPSITE ER Operating Manual
8.7.1.2 Additional Features of the Background Tool
By placing the mouse over a Background Subtract triangle and right-clicking, see
Figure 8-84, the following menu will be displayed:
Figure 8-84 Background Subtract Menu
Remove . . . . . . . . . . . . . . . . . . . . . . Removes the Background Subtract
triangle.
Select Mass for Integration . . . . . . Selects either the TIC or a mass fragment for
integration.
Show Integration . . . . . . . . . . . . . . Displays the integration on the x-axis.
Show Retention Time. . . . . . . . . . . Displays the retention time on the x-axis.
8.7.2 Range Tool
The Range Tool is a red-striped triangle located at the bottom left side of the
chromatogram. It is used to average spectra over a “range” of scans across a given
peak, especially when the analytes are low in concentration. It can also be used to
select a section of a peak or reintegrate peaks. The SI numbers for the selected
compound will increase when using the Range Tool. Follow the instructions below
to use the Range Tool.
1 Place cursor on the red-striped triangle, which is the Range Tool. Left-click,
hold and drag the triangle to the location where the scans should be averaged.
(See Figure 8-85.)
Figure 8-85 Range Tool