9 - 18
HAPSITE ER Operating Manual
Mass Calibration. . . . . . . . . . . . . . . Verifies and corrects the ten calibration
masses for correct location within the mass
range.
Noise Check . . . . . . . . . . . . . . . . . . Scans a quiet mass range (no peaks) to
determine the electronic baseline and
threshold noise level.
Perform Tune Checkup . . . . . . . . . Runs a mass calibration and noise check.
Save Tune Parameters... . . . . . . . . Saves the tune file.
Load Tune Parameters... . . . . . . . . Loads a new tune file and restarts tuning.
Load Factory Defaults . . . . . . . . . . Loads the default tune settings from a factory
tune file. This is intended to provide a starting
point for tuning.
Common Scale . . . . . . . . . . . . . . . . Sets all of the mass peak windows to the
same common scale (Y-axis), based on
Mass 117.
Show Tune Status Panel . . . . . . . . Displays the Tune and Mass Calibration
Status panel.
Show Mass Calibration Status . . . Displays the Mass Calibration Status
panel.
View Tune Reports . . . . . . . . . . . . . Displays the Tune Reports screen.
Properties . . . . . . . . . . . . . . . . . . . . Displays the Properties window, which is
used to set the default screen display and
startup/exit conditions for Manual Tune. See
Figure 9-23.
Advanced . . . . . . . . . . . . . . . . . . . . Displays the Advanced tune functions.
Linearize DACS. . . . . . . . . . . . . Repositions the mass peaks from the internal
standard gas on the mass axis by linear
extrapolation of the digital to analog control
settings.
AutoTune Tolerances . . . . . . . . Sets the AutoTune Tolerance for mass
resolution and mass axis position.
NOTE: The Advanced tune functions should only be utilized under the direction
of INFICON Support personnel.