9 - 22
HAPSITE ER Operating Manual
Running Tune Base Peak Gain . . . Current measured Base Peak Gain (BPG).
NOTE: The Base Peak Gain will switch to
red when BPG is outside the target
range.
Auto Resolve . . . . . . . . . . . . . . . . . Adjusts the resolution of all mass peaks to
the target resolution.
Save Tune . . . . . . . . . . . . . . . . . . . . Save the tune file.
Mass Cal.. . . . . . . . . . . . . . . . . . . . . Verifies and corrects the ten calibration
masses for correct location within the mass
range.
Full Scan . . . . . . . . . . . . . . . . . . . . . Switches between full scan display mode
and peak scan display mode
Short AutoTune. . . . . . . . . . . . . . . . Starts the AutoTune function
Noise Check . . . . . . . . . . . . . . . . . . Scans a quiet mass range (no peaks) to
determine the electronic baseline and
threshold noise level.
Tune Checkup . . . . . . . . . . . . . . . . . Runs a mass calibration and noise check.
Zoom . . . . . . . . . . . . . . . . . . . . . . . . Enables the cursor to zoom into full scan or a
section of the screen.
Mass Adjust . . . . . . . . . . . . . . . . . . Enables the cursor to select and drag a mass
peak to a different position on the mass axis.