10 - 27
HAPSITE ER Operating Manual
Figure 10-31 Method Editor SIM Page for Analyze Methods
The Scan Sets fields, in the order recommended for editing, are as follows:
Begin Time . . . . . . . . . . . . . . . . . . . The start time for mass collection.
End Time . . . . . . . . . . . . . . . . . . . . . The stop time for mass collection.
Name . . . . . . . . . . . . . . . . . . . . . . . . Each scan set can be assigned a name for
identification purposes. This entry is optional.
NOTE: One of the column entries listed above must be highlighted to enable
editing of the Mass list for that specific Scan Set.
Mass . . . . . . . . . . . . . . . . . . . . . . . . The mass fragments of each column are
entered in this column.
Mass Width . . . . . . . . . . . . . . . . . . . The width, in tenths of an amu, around the
mass which the Mass Spectrometer will
scan. For example, a Mass Width of 0.6 will
scan 0.3 amu on each side of the peak.