5 - 5
PN 074-550-P1D
SQC-310 Operating Manual
Controls on the Data Log window are:
None . . . . . . . . . . . . . . . . . . . . . . . . Data is not logged.
Overwrite. . . . . . . . . . . . . . . . . . . . . The log file will be overwritten each time data
logging is started with the Start Process or
Start Layer button on the Main window.
Append . . . . . . . . . . . . . . . . . . . . . . Data is added to the end of the log file each
time data logging is started with the Start
Process or Start Layer button on the Main
window.
New . . . . . . . . . . . . . . . . . . . . . . . . . Displays a file open dialog box, to allow
selection or creation of a new log file.
View . . . . . . . . . . . . . . . . . . . . . . . . . Displays the current log file using the default
Windows text file viewer.
Start Time . . . . . . . . . . . . . . . . . . . . The elapsed time when data logging begins.
The right value is seconds and the left value
is minutes.
End Time . . . . . . . . . . . . . . . . . . . . . The elapsed time when data logging ends.
The right value is seconds and the left value
is minutes.
Log Interval . . . . . . . . . . . . . . . . . . . The elapsed time between data log entries,
in seconds. Data will be logged as quickly as
possible when an interval value of zero is
entered. This time will vary with system
configuration.
Once communications have been established, follow these steps to start data
acquisition:
1 Select the desired process/layer from the list of SQC-310 processes/layers.
2 Press Zero to zero the thickness reading (optional).
3 With the log file option set to append or overwrite, Press Start Process or Start
Layer to start data acquisition.