3 INSTALLATION DACS-W Operation Manual
3-2
3.2 Installation Location & Environment
The weigher installation site should conform to the following conditions:
Indoors
Ambient temperature range: 0-40 °C
Ambient Humidity: 35-85% (no condensation)
Installed on a hard horizontal surface with minimum vibration
Sufficient area around weigher to perform maintenance
No electrical interference from such devices as a wireless transmitter
3.3 Power Source Connection
All on-site electrical work should be performed by qualified electrical contractors.
Do not connect power devices which may emit electrical interference, e.g. a
motor, to the same power line as the weigher.
Voltages should not vary more than ±
±±
±10% under load.
Two types of DACS-W electrical specifications are available: AC100V and AC200V.
Check the rating plate attached to the side of the RCU to confirm specification type.
Use a power outlet that conforms to the specified voltage.
Be sure that a ground wire is connected to the outlet. Power cables from the checkweigher should be
connected by an Ishida Service representative.