5.3 Explanation of Each Program Message
BP4610
5-12
SEXP/SEXP?
Description: Sets the step execution parameters.
Parameter: 1st parameter: Sets the DC voltage/DC current.
The DC voltage can be set in units of 0.01 V in the range of 115.00 V to
115.00 V.
The DC current can be set in units of 0.001 A in the range of 10.000A to
10.000A.
2nd parameter: Sets the DC voltage/DC current operation type.
0: CONST/1: KEEP/2: SWEEP
3rd parameter: Sets the superimposed AC voltage/superimposed AC current.
The AC voltage can be set in units of 0.1 Vp-p in the range of 0.0 Vp-p to
120.0 Vp-p.
The AC current can be set in units of 0.01 Ap-p in the range of 0.00 Ap-p
to 30.00 Ap-p.
4th parameter: Sets the superimposed AC voltage/superimposed AC current
operation type.
0: CONST/1: KEEP/2: SWEEP
5th parameter: Sets the frequency.
The frequency can be set in units of 0.1 Hz in the range of 1.0 Hz to
100000.0 Hz.
6th parameter: Sets the frequency operation type.
0: CONST/1: KEEP/2: SWEEP
7th parameter: Sets the waveform.
2 to 17: ARB1 to ARB16
8th parameter: Sets the waveform operation type.
0: CONST/1: KEEP
9th parameter: Sets the step sync output.
0: (0,0)/1: (0,1)/2: (1,0)/3: (1,1)
10th parameter: Sets the step sync output operation type.
0: CONST/1: KEEP
Setting example: Setting of DC voltage to 10 V, SWEEP, AC voltage to 5 Vp-p superimposed,
KEEP, frequency to 20 Hz, KEEP, waveform to ARB3, CONST, step sync
output to (0,0), KEEP
SEXP 10, 2, 5, 1, 20, 1, 4, 0, 0, 1
SLEN?
Description: Specifies the sequence length.
Parameter: 255, fixed
SSTS?
Description: Returns the sequence status.
Response format: 0: Idle status
1: Run status
2: Hold status