z With types of dyeing of the specimen to observe, the Barrier Filters will be set
automatically to the optical path.
To change the Barrier Filter types, see section 1-3-2-6,”Configuring the Filters “ and
change by [Optical system Configuration] window.
z Turn off the transmitted light bulb of the microscope.
1. From the page tabs on the bottom right of the [Acquire] panel, select the [Settings]
sub-panel.
Fig. 2-23 [Settings] Sub-panel
2. Click the <Tr. Lamp> button to turn the lamp off (set the button to the non-pressed
status).
z In the [Acquire] panel, make sure that the [Transmitted] check box in the [Ch5] group
box is check-marked to indicate that Ch3 is ready for image acquisition.
<Tr. Lamp> button
Turns off the transmitted light bulb of the
microscope.
(The lamp lights when the button is in
the pressed status.)
Ch5 check box