EasyManua.ls Logo

Stratasys Objet30 - Page 25

Stratasys Objet30
237 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Objet30 User Guide 2 Safety
DOC-34025 Rev. I 2-10
UV Lamps
UV lamps used by the printer to cure printing materials contain a small amount
of mercury, and are considered “Universal Waste.” Recycle or discard used
lamps in accordance with applicable regulations.
Broken lamps:
After ventilating the area, use protective gloves and carefully remove spilled
mercury with a method that prevents the generation of mercury vapor, such as
a syringe, packing tape or paper. Place the broken lamp, mercury and
contaminated materials in an air-tight, non-metallic container. Dispose of the
container in accordance with applicable regulations.