<Chromogenic Method>
Low Light Error Limit: A/D limit to detect low transmitted light intensity.
High Light Error Limit: A/D limit to detect high transmitted light intensity.
Reagent Correlation a: Coefficient of delta-OD compensation.
Reagent Correlation b: Offset of delta-OD compensation
The delta-OD after compensation = a x (Detected delta-OD) + b
VlinIntegral Algorithm: Use of VlinIntegral Algorithm
Polynomial Dimension: Dimension of polynomial approximate expression
Start Time: The start time for analysis.
End Time: The end time for analysis.
Integral Area(auc): The judgement value of polynomial approximate expression result
Minimum Search Window: Width to search variable analysis starting time
Variable Start Time: Use of variable analysis starting time
Search Width for Preevaluation: The modulus for variable analysis starting time calculation
Upper max Evaluation Offset: The modulus for variable analysis starting time calculation
Upper Preevaluation Rate: The modulus for variable analysis starting time calculation
Lower Preevaluation Rate: The modulus for variable analysis starting time calculation
Min. Regression Time: The modulus for linearity range
Mixing Time for Sample: The mixing time for sample
Mixing Time for Reagent: The mixing time for reagent
Calibration a: The slant of calibration
Calibration b: The segment of calibration
High Accurate Dilute ratio The parameter for dilute ratio of standard curve analysis.
ON: accuracy equal to the CA-1500 OFF: conventional accuracy 11
<Immunoassy Method>
Low Light Error Limit: A/D limit to detect low transmitted light intensity.
High Light Error Limit: A/D limit to detect high transmitted light intensity.
Reagent Correlation a: Coefficient of delta-OD compensation.
Reagent Correlation b: Offset of delta-OD compensation
The delta-OD after compensation = a x (Detected delta-OD) + b
Start Time a1: The start time for Prozone check a1.
Stop Time a1: The end time for Prozone check a1.
Start Time a2: The start time for Prozone check a2.
Stop Time a2: The end time for Prozone check a2.
Cut Off: A parameter for Prozone check a1.
Rate: A parameter for Prozone check.
Offset: A parameter for Prozone check.
VlinIntegral Algorithm: Use of VlinIntegral Algorithm
Polynomial Dimension: Dimension of polynomial approximate expression
Start Time: The start time for analysis.
End Time: The end time for analysis.
Integral Area(auc): The judgement value of polynomial approximate expression result
Minimum Search Window: Width to search variable analysis starting time
Variable Start Time: Use of variable analysis starting time
Search Width for Preevaluation: The modulus for variable analysis starting time calculation
Upper max Evaluation Offset: The modulus for variable analysis starting time calculation
Upper Preevaluation Rate: The modulus for variable analysis starting time calculation
Lower Preevaluation Rate: The modulus for variable analysis starting time calculation
Min. Regression Time: The modulus for linearity range
Mixing Time for Sample: The mixing time for sample
Mixing Time for Reagent: The mixing time for reagent
Calibration a: The slant of calibration
Calibration b: The segment of calibration
High Accurate Dilute ratio The parameter for dilute ratio of standard curve analysis.
ON: accuracy equal to the CA-1500 OFF: conventional accuracy 11
[Print] Key
Use to output the settings for all tests to the built-in printer.
Numeric Keys
When the cursor is moved to each setting value, numeric keys appear as below to input the value.
CA-500 Series S/M 5-15 Revised December 2001 8