(7) Check the Sensor Status and Locking Function
1) Turn ON the power.
2) Change the Barcode Scanner setting to “connected” Refer to Operator’s Manual
Chapter 10 5. Devices to be connected for more detail.
3) Select [Sensor Status] by selecting [Special Menu] Æ [Special Operate] Æ [System
Tests] Æ [System Test] Æ [Sensor Status].
4) Make sure that [Sample Table] status changes from ““ to ““ when Sampler position
changes from in to out. (“IN” position : ““ ,”OUT” position : ““ )
5) Make sure that Sampler cannot be pull out when [Lock] key is pressed on Sampler’s IN
position. Make sure that [Lock] key changes to [Unlock] when [Lock] key is pressed.
6) Make sure that Sample can be pulled out when [Unlock] key is pressed.
(8) Check the Reading Function
1) Select [Sensor Status] by selecting [Special Menu] -> [Special Operate] -> [System
Tests] -> [System Test] -> [Barcode Scanner].
2) Make sure that bar code scanner moves to the home position when [Ret ORG] key is
pressed. Make sure that [ORG Sensor] status changes from ““ to “ when the bar
code scanner reaches to the home position. (There is a case that [ORG Sensor]
status changes from“ to ““ and changes to ““ in a very short time. This is also
normal function.)
3) Set 10 sample test tubes with bar code label in the sample rack.
4) Perform reading test by pressing [CONTINUE] key. (Refer to Operator’s Manual
Chapter 8 5. Barcode Scanner.)
5) Make sure that all labels are read without an error.
(9) Adjust the Pipette Position for the Sampler
1) Perform pipette position adjustment. Refer to Service Manual 4.3.4 Position
Adjustment Procedure of Pipette (“(1)blood 1”, “(2)blood 10” and “(3)blood stat”).
CA-500 Series S/M B-18 Revised December 2001 8