Quadra 4 Operator Manual
Tomtec
168
12) With Proper Pipette Tips on Stacker Nest, Raise stage to Safe Height then the up
arrow until the tips are firmly seated on the Tip Pins without back driving the stage
motor.
13) With tips firmly seated – Select Record Load Height. This will record the proper
stage height
14) Select Safe Height again – this will be the tip shuck (remove) height
15) Select Record Shuck Height
16) Select Set Load Height
17) Select Set Shuck Height
18) In Test Load – Shuck Tips function, select the tip volume for the pipette tips that
have just been loaded
19) Initialize Stage
20) Select OK
21) Return to Tip Utility
22) Select System Utilities
23) Select Shuck Tips
24) Place another set of pipette tips (different volume) on the nest
25) Repeat procedure for other tip volumes