USTER®
HVI 900
Instruction Manual
Uster Technologies, Inc.
Chapter 5
Version 4.0 (1.99) Calibration - Page 5-28
6.6 Rd and +b Tolerance
• To access the RD AND +B TOLERANCE MENU from the CALIBRATION
MENU, highlight the item using the arrow keys, then press ENTER.
This item is used to enter the tolerance values for the Rd and +b. These
values are used to determine the Pass/Fail during color calibration.
•
Enter the Rd tolerance value for each tile color by pressing ENTER,
then typing the value. Press ENTER again to enter the +b tolerance
value. When the tolerance values are correct, press ESCAPE to
return to the C
OLOR CALIBRATION MENU.