3.5 Length Standard Calibration
Menu
Use the arrow keys to highlight the LENGTH STANDARD CALIBRATION
MENU item from the SYNTHETIC FIBER MENU, then press ENTER.
Note: The metal length standard must always be entered in inches
even though the fiber measurements may be made in millimeters.
The metal length standard is used to check and update the accuracy of
the value that measures the distance from the comb transport’s home
position to the center of the optics.
Only use the metal length standard that is provided with the instrument
for this calibration. If the metal length standard is changed, a
corresponding change must be made in the values entered in the
software.