Chapter 7
Version 4.0 (1.99) Module Testing - Page 7-11
The Optics and Force values, as well as the Temperature and Relative
Humidity values, are continuously updated and displayed at the bottom
of the screen.
3.6 Length/Strength Module Testing
Procedures
• If test results will be sent to a printer or transmitted to a host
computer, an ID must be entered on the test screen. If the data will
not be printed or transmitted, no ID is requested.
• If a micronaire value is required (via the keyboard or micronaire),
"Enter MIC:" is displayed on the screen. Enter a micronaire value
between 2.0 and 8.0, and press ENTER. If the micronaire value
has been established, use that value; if it is unknown, use the
standard value of 4.0.
If the micronaire value is to be determined via the micronaire test,
"Mic" from 8.5 to 11.5 grams of fiber. Mass:" is displayed. Weigh
the sample on the balance and then place the sample in the
chamber for the test.
If the Micronaire Value is set to
None
, the Operations options (listed
below) are displayed immediately after the ID has been entered. If
None is selected, the system default value of 4.0 is used.
• The following Operation options will display after the ID and
Micronaire value (if required) have been entered.
— Press ESCAPE to exit
—
Press <!> to reject (To reject the last test)
— Press <#> to set motors