USTER®
HVI 900
Instruction Manual
Uster Technologies, Inc.
Chapter 2
Version 4.0 (1.99) Installation - Page 2-3
The instrument must be operated in a controlled environment in the
range of 20°C ±2°C (65°F to 72°F). The relative humidity should be
65%, ±2%. For consistent results, fiber samples should be conditioned
in the laboratory environment.
4 Compressed Air
The system requires 90-125 psi (620-825 kPa) of clean, dry,
unregulated compressed air. The unit is provided with an air line for
connections to the facility's air supply. Failure to supply air that is free
of water, oil, or particles will increase maintenance and void the
warranty.
Attention: Damage caused by connecting the instrument
to improper power and air sources is not covered under
the warranty.