USTER®
HVI 900
Instruction Manual
Uster Technologies, Inc.
Chapter 5
Version 4.0 (1.99) Calibration - Page 5-36
8.5 Set Current Value
An entry box is displayed to enter the current temperature. When this
selection is used, only one point is used to adjust the slope from the
current value. The slope must be reasonably correct when this selection
is used.
8.6 Show Current Value
This selection displays the current temperature.
8.7 Temperature Slope and
Temperature Offset
These values are automatically calculated and updated based on the
information entered for calibration. Both Slope and Offset values are
updated when Trend Calibration is used. Only the Slope is changed
when Set Current Value is the basis.