EasyManua.ls Logo

USTER HVI 900 - Starting Sample Code and Number of Samples

USTER HVI 900
340 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
USTER®
HVI 900
Instruction Manual
Uster Technologies, Inc.
Chapter 5
Version 4.0 (1.99) Calibration - Page 5-30
7.5 Starting Sample Code and
Number of Samples
The number entered for Number of Samples determines how many
samples the software will request to test during Sample Calibration. The
Starting Sample Code should be the code for the first sample to be
viewed during Sample Calibration. After the Starting Sample Code is
entered, the software will assign sample codes to the remaining
samples.
Use the calibration samples that are supplied with the HVI instrument.
Enter the values by pressing ENTER, then typing the value. Press
ENTER again to enter next value. Once the values are correct,
press ESCAPE to save the data.