Version 4.0 (1.99) System Maintenance - Page 12-2
that the top of the cabinet can be lifted.
Lift the top of the Length/Strength cabinet, and vacuum all areas
under the cover. Pay particular attention to cleaning the comb track
area.
After cleaning, close all doors and covers to their original positions.
3.6 Daily
The Lint Waste Box on Version 3 instruments should be cleaned daily.
The vacuum on the box is relieved using the vacuum ON/OFF switch
on the front panel, by opening the instrument doors, or by raising the
top. Use the handles to open the lint/waste box doors and remove
all fibers from the lint box by hand. Clean the filter inside the waste
box with a brush (or vacuum cleaner) removing all loose material.
Upon completion, close the door. Both doors must be closed for
proper operation.
Using a clean, dry cloth; wipe off the two rails on which the comb
assembly moves left and right.
Clean the Color Window.
Inspect the computer's Cooling Fan Filters, and clean if necessary.
3.7 Weekly
Check the Inlet/Filter air filters at least once a week to see if water has
collected in the filter. If so, remove the water by pushing the release
valve on the bottom of the glass bulb.