Chapter 3
Version 4.0 (1.99) System Startup - Page 3-7
4.10 Shutdown
This item should be used to leave the system on, but not in use, for long
periods of time. It clears the monitor screens, leaving the message,
"Press any key to restart." When any key on the keyboard is pressed,
the monitor is reactivated and the M
AIN MENU is displayed.
It is recommended that at the end of each day of testing the printer and
monitors be turned OFF but the instrument be left ON. This is because if
the instrument's main power switch has been OFF, at least 4 hours must
be allowed for the colorimeter to warm up before calibrating and testing
can begin.
4.11 Quit
This selection exits the HVI system software and goes to the DOS
prompt.
5 Resetting the HVI 900 System
The main purpose of resetting the HVI 900 system is to clear the
software of unnecessary information. Resetting erases old values from
the system's memory buffer. If the system is reset in the middle of a
testing sequence, the result will be a loss of test data.
The system can also be reset if it is not responding to keyboard strokes.
Rest the instrument (reload the program) as follows while in any part of
operation with the exception of
Testing
.
Open the front doors of the Length/Strength cabinet, and press the
button labeled RESET.