Chapter 5
Version 4.0 (1.99) Calibration - Page 5-35
Enter the number of trend points to be used (between 2 and 10).
The screen provides entry points for the desired and current value
for each of the points. As each value is entered, the cursor moves
to the next entry. To correct erroneous values that are entered on
the Trend Calibration Screen, press the BACKSPACE key before
pressing ENTER.
To exit the Trend Calibration screen and return to the TEMPERATURE
CALIBRATION MENU without altering the current Slope and Offset
values, press ESCAPE. A message is displayed on the screen that
reads "Abort?" Enter "Y" to return to the Temperature Calibration
screen; the values entered for Trend Calibration are ignored. If you
press "N" at the "Abort" message, the Trend Calibration screen
remains, and the cursor goes to the next field for the value to be
entered.
Once the cursor has moved to the next calibration point, the entry
cannot be corrected. You must exit the Trend Calibration screen by
entering "Y" at the "Abort" prompt, and then select
Trend Calibration
again.