USTER®
HVI 900
Instruction Manual
Uster Technologies, Inc.
Chapter 5
Version 4.0 (1.99) Calibration - Page 5-5
• Remove your hands and other objects from the area and press
ENTER to begin the test. The instrument will perform the test using
the length standard and it will display the results indicating whether
the value measured Passed or Failed.
• After the test is complete, remove the metal length standard by
pushing it from the left side of the track the right side using your
fingers or another narrow object such as another fibrocomb until is
can be pulled from the right side and removed.
If the metal length standard will not slide in/out from the track easily,
refer to the following instructions to raise the cabinet top and
insert/remove the comb.
• Swing the monitor to the side so that the cabinet top can be lifted.
Make sure no loose objects are lying on the cabinet top. Lift the
cabinet top, making sure that the brace on the rod spring lock is
engaged when the cabinet top is raised. The blower will
automatically shut off when the top is lifted.
• Slide the length standard along the track into the comb holder, and
position it in front of the clamping mechanism.
• Flip the brace from the locked position, and lower the cabinet top to
its original position. The blower will automatically turn on after the
top is closed.
• Slide the length standard along the track into the comb holder, and
position it in front of the clamping mechanism.
•
Press ENTER to begin the test. The instrument will perform the test
using the length standard and then it will display the results
indicating whether the value measured Passed or Failed.