USTER®
HVI 900
Instruction Manual
Uster Technologies, Inc.
Chapter 5
Version 4.0 (1.99) Calibration - Page 5-12
If all parameters are within the tolerance levels (passed), the instrument
is properly calibrated. The LENGTH/STRENGTH CALIBRATION MENU will
reappear when the ENTER key is pressed
4.6.3 Cotton Calibration Test Results—Fail
Condition
If there is a difference of one or more between the Standard Values and
Average measured values that exceeds the Tolerance values, a "Fail"
condition will result.
The "Fail" condition may be a result of one or more of the following
situations.
♦ The calibration cottons may have been mixed or contaminated
with other cottons by mistake.
♦ Calibration cotton moisture content may have changed since the
previous calibration.