USTER®
HVI 900
Instruction Manual
Uster Technologies, Inc.
Chapter 5
Version 4.0 (1.99) Calibration - Page 5-12
If all parameters are within the tolerance levels (passed), the instrument
is properly calibrated. The LENGTH/STRENGTH CALIBRATION MENU will
reappear when the ENTER key is pressed
4.6.3 Cotton Calibration Test Results—Fail
Condition
If there is a difference of one or more between the Standard Values and
Average measured values that exceeds the Tolerance values, a "Fail"
condition will result.
The "Fail" condition may be a result of one or more of the following
situations.
The calibration cottons may have been mixed or contaminated
with other cottons by mistake.
Calibration cotton moisture content may have changed since the
previous calibration.