the tubehead connector X56 is well inserted and then perform the tests of
the Error E761 (see paragraphs 9.2.7.1 and 9.2.7.9).
8.
If the problem is on the ceph sensor check the ethernet cable from the ceph arm
to the ethernet switch and the ethernet cable to the PC. Check also if the detector
is well inserted in its holder. Try to move the sensor and check that the blue led
(sensor power supply) is always ON
Contact Technical Service providing the following additional information:
-
Software logs
-
Ceph Sensor logs
-
3D Sensor logs folder(see paragraph 11.2.1)
-
Last RAW files folder stored (see paragraph 11.2.1.5)