Agilent N5181A/82A MXG Signal Generators SCPI Command Reference 95
System Commands
Status Subsystem (:STATus)
The variable <value> is the sum of the decimal values of the bits that you want to enable.
Range 0–32767
Remarks Refer to the Programming Guide for more information.
:QUEStionable:CALibration[:EVENt]
Supported All
CAUTION This is a destructive read. The data in the register is latched until it is queried. Once
queried, the data is cleared.
:STATus:QUEStionable:CALibration[:EVENt]?
This command returns the decimal sum of the bits in the Data Questionable Calibration Event
Register.
Range 0–32767
Remarks The equivalent PTR or NTR filters must be set before the condition register can
set the corresponding bit in the event register.
Refer to the Programming Guide for more information.
:QUEStionable:CONDition
Supported All
:STATus:QUEStionable:CONDition?
This query returns the decimal sum of the bits in the Data Questionable Condition Register. For
example, if the ALC Heater Detector is cold (bit 4), a value of 16 is returned.
Range 0–32767
Remarks The data in this register is continuously updated and reflects current conditions.
Refer to the Programming Guide for more information.
:QUEStionable:ENABle
Supported All
:STATus:QUEStionable:ENABle <value>
:STATus:QUEStionable:ENABle?
This command determines which bits in the Data Questionable Event Register will set the Data
Questionable Status Group Summary bit (bit 3) in the Status Byte Register.
The variable <value> is the sum of the decimal values of the bits that you want to enable.
Range 0–32767
Remarks Refer to the Programming Guide for more information.