126 Agilent N5181A/82A MXG Signal Generators SCPI Command Reference
Analog Modulation Commands
Frequency Modulation Subsystem–Option UNT ([:SOURce])
Key Entry FM Off On
Remarks The RF carrier is modulated when you set the signal generator’s modulation state
to ON, see “:MODulation[:STATe]” on page 89 for more information.
Whenever frequency modulation is enabled, the FM annunciator is turned on in the
display.
:FM[:DEViation]
Supported All Models with Option UNT
[:SOURce]:FM[:DEViation] <value><unit>
[:SOURce]:FM[:DEViation]?
This command sets the frequency modulation deviation.
*RST +1.00000000E+003
Key Entry FM DEV
:FM[:DEViation]:STEP[:INCRement]
Supported All Models with Option UNT
[:SOURce]:FM[:DEViation]:STEP[:INCRement] <value><unit>|GHz|MHz|kHz|Hz
[:SOURce]:FM[:DEViation]:STEP[:INCRement]?
This command sets the step increment for the FM deviation of the signal generator.
*RST +5.00000000E+003
Key Entry Incr Set
Remarks The setting enabled by this command is not affected by signal generator power- on,
preset, or *RST.
Range Frequency Deviation
<250 MHz
a
a.Settable, but not specified to 100 kHz.
0–20 MHz
250 MHz – <375 MHz 0–2.5 MHz
375 MHz – <750 MHz 0–5 MHz
750 MHz – <1.5 GHz 0–10 MHz
1.5 GHz – <3.000001 GHz 0–20 MHz
3.000001 GHz – 6 GHz 0–40 MHz