26 Agilent N5181A/82A MXG Signal Generators SCPI Command Reference
Basic Function Commands
Digital Modulation Subsystem—N5182A ([:SOURce])
ON. Refer to “:DM:IQADjustment[:STATe]” on page 30.
:DM:IQADjustment:EXTernal:DIOFfset
Supported N5182A
[:SOURce]:DM:IQADjustment:EXTernal:DIOFfset <value>
[:SOURce]:DM:IQADjustment:EXTernal:DIOFfset?
This command sets the differential offset voltage for an in- phase (I) signal routed to the I output
connectors.
The variable <value> is expressed in units of volts (mV–V).
*RST +0.00000000E+000
Range −25mV to 25mV
Key Entry Diff. Mode I Offset
Remarks This command is effective only if the state of the I/Q adjustment function is set to
ON. Refer to “:DM:IQADjustment[:STATe]” on page 30.
:DM:IQADjustment:EXTernal:DQOFfset
Supported N5182A
[:SOURce]:DM:IQADjustment:EXTernal:DQOFfset <value>
[:SOURce]:DM:IQADjustment:EXTernal:DQOFfset?
This command sets the differential offset voltage for a quadrature- phase (Q) signal routed to the Q
output connectors.
*RST +0.00000000E+000
Range −25mV to 25mV
Key Entry Diff. Mode Q Offset
Remarks This command is effective only if the state of the I/Q adjustment function is set to
ON. Refer to “:DM:IQADjustment[:STATe]” on page 30.
:DM:IQADjustment:EXTernal:IOFFset
Supported E4438C
[:SOURce]:DM:IQADjustment:EXTernal:IOFFset <value>
[:SOURce]:DM:IQADjustment:EXTernal:IOFFset?
This command sets the offset voltage for a signal applied to the 600 ohm I input connector.
The variable <value> is expressed in units of volts (mV–V).
*RST +0.00000000E+000
Key Entry External Input I Offset
Range −100 mV to 100 mV