EasyManua.ls Logo

Agilent Technologies n5181a - Correction Subsystem ([:Source]:Correction); Flatness:initialize:fstep; Flatness:load

Agilent Technologies n5181a
270 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
20 Agilent N5181A/82A MXG Signal Generators SCPI Command Reference
Basic Function Commands
Correction Subsystem ([:SOURce]:CORRection)
Correction Subsystem ([:SOURce]:CORRection)
:FLATness:INITialize:FSTep
Supported All Models
CAUTION The current flatness data will be overwritten once this command is executed. If needed,
save the current data. Refer to “:FLATness:STORe” on page 23 for storing user flatness
files.
[:SOURce]:CORRection:FLATness:INITialize:FSTep
This command replaces the loaded user flatness data with the settings from the current step array
data points.
You can load only one user flatness file at a time.
The maximum number of user flatness points is 1,601. When copying the step array settings over to
a user flatness file, ensure that the number of points in the step array do not exceed the maximum
user flatness points.
Key Entry Load Cal Array From Step Array
:FLATness:LOAD
Supported All Models
[:SOURce]:CORRection:FLATness:LOAD "<file name>"
This command loads a user- flatness correction file. The "<file name>" variable is the name of the file
located in the Catalog of USERFLAT Files. The directory path is implied in the command and need
not be specified in the variable name. For more information on file name syntax, refer to “File Name
Variables on page 12.
Key Entry Load From Selected File

Table of Contents

Other manuals for Agilent Technologies n5181a

Related product manuals