70 Agilent N5181A/82A MXG Signal Generators SCPI Command Reference
System Commands
Diagnostic Subsystem (:DIAGnostic[:CPU]:INFOrmation)
:OTIMe
Supported All Models
:DIAGnostic[:CPU]:INFormation:OTIMe?
This query returns the cumulative number of hours that the signal generator has been on.
Key Entry Diagnostic Info
:REVision
Supported All Models
:DIAGnostic[:CPU]:INFormation:REVision?
This query returns the CPU bootstrap read only memory (boot ROM) revision date. In addition, the
query returns the revision, creation date, and creation time of the main firmware.
Key Entry Diagnostic Info
:SDATe
Supported All Models
:DIAGnostic[:CPU]:INFormation:SDATe?
This query returns the date and time of the firmware revision.
Key Entry Diagnostic Info
:WLICence[:VALue]
Supported N5182A with Option 651/652/654
:DIAGnostic[:CPU]:INFormation:WLIcense[:VALue]? <type_num>
This query returns the number of seconds remaining on the waveform license for the type of
waveform designated by the <type_num> variable number. The type variable number is obtained using
the :LICense:WAVeform command shown on page 69. Zero is returned for non- existent and expired
licenses. The value 2^32 - 1 (4,294,967,295) is returned for licenses that do not expire.