Agilent N5181A/82A MXG Signal Generators SCPI Command Reference 47
Basic Function Commands
List/Sweep Subsystem ([:SOURce])
:LIST:TYPE
Supported All Models
[:SOURce]:LIST:TYPE LIST|STEP
[:SOURce]:LIST:TYPE?
This command toggles between the two types of sweep.
LIST This type of sweep has arbitrary frequencies and amplitudes.
STEP This type of sweep has equally spaced frequencies and amplitudes.
*RST STEP
Key Entry Sweep Type List Step
:LIST:TYPE:LIST:INITialize:FSTep
Supported All Models
CAUTION The current list sweep data will be overwritten once this command is executed. If
needed, save the current data. Refer to “:STORe:LIST” on page 85 for storing list sweep
files.
[:SOURce]:LIST:TYPE:LIST:INITialize:FSTep
This command replaces the loaded list sweep data with the settings from the current step sweep data
points.
You can load only one sweep list at a time.
The maximum number of list sweep points is 1,601. When copying the step sweep settings over to a
list sweep, ensure that the number of points in the step sweep do not exceed the maximum list
sweep points.
Key Entry Load List From Step Sweep
:LIST:TYPE:LIST:INITialize:PRESet
Supported All Models
CAUTION The current list sweep data will be overwritten once this command is executed. If
needed, save the current data. Refer to “:STORe:LIST” on page 85 for storing list sweep
files.
[:SOURce]:LIST:TYPE:LIST:INITialize:PRESet
This command replaces the current list sweep data with a factory- defined file consisting of one point
at a frequency, amplitude, and dwell time.
Key Entry Preset List