Agilent N5181A/82A MXG Signal Generators SCPI Command Reference 41
Basic Function Commands
Frequency Subsystem ([:SOURce])
:ROSCillator:SOURce
Supported All Models
[:SOURce]:ROSCillator:SOURce?
This command queries the current reference oscillator source: INT (internal) or EXT (external).
:ROSCillator:SOURce:AUTO
Supported All Models
[:SOURce]:ROSCillator:SOURce:AUTO ON|OFF|1|0
[:SOURce]:ROSCillator:SOURce:AUTO?
This command enables or disables the ability of the signal generator to automatically select between
the internal and an external reference oscillator.
ON (1) This choice enables the signal generator to detect when a valid reference signal is
present at the 10 MHz IN connector and automatically switches from internal to
external frequency reference.
OFF (0) This choice selects the internal reference oscillator and disables the switching
capability between the internal and an external frequency reference.
*RST 1
Key Entry Ref Oscillator Source Auto Off On