Agilent N5181A/82A MXG Signal Generators SCPI Command Reference 123
Analog Modulation Commands
Amplitude Modulation Subsystem–Option UNT ([:SOURce])
Remarks The value set by this command is used with the UP and DOWN choices for the
AM depth setting. Refer to “:AM[:DEPTh][:LINear]” on page 123 for more
information.
The setting enabled by this command is not affected by signal generator power- on,
preset, or *RST.
:AM[:DEPTh][:LINear]
Supported All Models with Option UNT
[:SOURce]:AM[:DEPTh][:LINear] <value><unit>|UP|DOWN
[:SOURce]:AM[:DEPTh][:LINear]?
This commands sets the amplitude modulation depth in percent.
*RST +1.00000000E001
Range 0.00–90%
Key Entry AM Depth
Remarks Refer to “:AM[:DEPTh]:STEP[:INCRement]” on page 122 for setting the value
associated with UP and DOWN choices.