24 Agilent N5181A/82A MXG Signal Generators SCPI Command Reference
Basic Function Commands
Digital Modulation Subsystem—N5182A ([:SOURce])
Digital Modulation Subsystem—N5182A ([:SOURce])
:BURSt:STATe
Supported N5182A
[:SOURce]:BURSt:STATe ON|OFF|1|0
[:SOURce]:BURSt:STATe?
This command enables or disables the burst envelope function.
*RST 0
Key Entry Burst Envelope Off On
:DM:CORRection:OPTimization
Supported N5182A
[:SOURce]:DM:CORRection:OPTimization RFOut|EXTernal
[:SOURce]:DM:CORRection:OPTimization?
This command enables the internal optimized path to accommodate I/Q signals.
EXT This choice enables the Baseband Generator to external I/Q output path, applies
correction terms, and provides a calibrated signal at the IQ output. When the
I/Q Output is selected, the RF signals at the RF Output are uncalibrated.
RFO This choice enables the Baseband Generator to RF output path, applies correction
terms, and provides a calibrated signal at the RF output. When the RF Output is
selected, the I/Q signals at the I/Q Output are uncalibrated.
*RST RFO
Key Entry I/Q Correction Optimized Path
:DM:EXTernal:ALC:BANDwidth|BWIDth
Supported N5182A
NOTE Refer to the Programming Compatibility Guide for information on this command. This
command was replaced by the “:ALC:BANDwidth|BWIDth” command shown in Chapter 6.