76 Agilent N5181A/82A MXG Signal Generators SCPI Command Reference
System Commands
IEEE 488.2 Common Commands
*OPC?
Supported All Models
*OPC?
The Operation Complete (OPC) query returns the ASCII character 1 in the Standard Event Status
Register when all pending operations have finished.
This query stops any new commands from being processed until the current processing is complete.
*OPT?
Supported All Models
*OPT?
The options (OPT) query returns a comma- separated list of all of the instrument options currently
installed on the signal generator.
Key Entry Instrument Options
*PSC
Supported
*PSC ON|OFF|1|0
The Power- On Status Clear (PSC) command controls the automatic power- on clearing of the Service
Request Enable Register, the Standard Event Status Enable Register, and device- specific event enable
registers.
ON (1) This choice enables the power- on clearing of the listed registers.
OFF (0) This choice disables the clearing of the listed registers and they retain their status
when a power- on condition occurs.
Remarks The setting enabled by this command is not affected by signal generator power- on,
preset, or *RST.
*PSC?
Supported All Models
*PSC?
The Power- On Status Clear (PSC) query returns the flag setting as enabled by the *PSC command.
*RCL
Supported All Models
*RCL <reg>,<seq>
The Recall (RCL) command recalls the state from the specified memory register <reg> of the specified
sequence <seq>.
Range Registers: 0–99 Sequences: 0–9