48 Agilent N5181A/82A MXG Signal Generators SCPI Command Reference
Basic Function Commands
List/Sweep Subsystem ([:SOURce])
:LIST:WAVeform
Supported N5182A
CAUTION The current list sweep data will be overwritten once this command is executed. If
needed, save the current data. Refer to “:STORe:LIST” on page 85 for storing list sweep
files.
[:SOURce]:LIST:WAVeform <name>{,<name>}
[:SOURce]:LISt:WAVeform?
This command sets the waveform values for the current list waveform sequence.
Example
:LIST:WAV "WFM1:RAMP_TEST_WFM","WFM1:SINE_TEST_WFM"
The preceding example loads the waveforms RAMP_TEST_WFM and SINE_TEST_WFM into the waveform
section of the List Table.
Remarks The setting enabled by this command is not affected by signal generator power- on,
preset, or *RST.
:LIST:WAVeform:POINts
Supported N5182A
[:SOURce]:LIST:WAVeform:POINts?
This query returns the number of waveform points in the current list sweep file.
:SWEep:DWELl
Supported All Models
[:SOURce]:SWEep:DWELl <value>
[:SOURce]:SWEep:DWELl?
This command enables you to set the dwell time for a step sweep.
The variable <value> is expressed in units of seconds with a 0.001 resolution.
The dwell time is the amount of time the sweep is guaranteed to pause after setting the frequency
and/or power for the current point.
NOTE The dwell time (<value>) does not begin until the signal generator has settled for the current
frequency and/or amplitude change.
*RST +2.00000000E003
Range 0.0001–100
Key Entry Step Dwell