Agilent N5181A/82A MXG Signal Generators SCPI Command Reference 117
System Commands
Trigger Subsystem
*RST IMM
Remarks The wait for the BUS, EXTernal, or KEY trigger can be bypassed by sending the
:TRIGger[:SEQuence][:IMMediate] command.
Example
:TRIG:SOUR BUS
The preceding example sets the sweep trigger source to BUS.
*RST IMM
:TRIGger[:SEQuence]:TIMer
Supported All Models
:TRIGger[:SEQuence]:TIMer <period>
:TRIGger[:SEQuence]:TIMer?
This command sets the period of the timer trigger.
*RST 1 ms
Range .5ms–1000s
Key Entry Trig Timer Period
:TRIGger[:SEQuence][:IMMediate]
Supported All Models
:TRIGger[:SEQuence][:IMMediate]
This event command causes an armed List or Step sweep to immediately start without the selected
trigger occurring.
Key Entry Bus Free Run Ext Trigger Key Timer Trigger