Agilent N5181A/82A MXG Signal Generators SCPI Command Reference 21
Basic Function Commands
Correction Subsystem ([:SOURce]:CORRection)
:FLATness:PAIR
Supported All Models
[:SOURce]:CORRection:FLATness:PAIR <freq.>[<freq suffix>],
<corr.>[<corr suffix>]
This command sets a frequency and amplitude correction pair.
The maximum number of points that can be entered is 1601. Option 501 is specific to the N5181A.
<corr.> This variable is the power correction.
Key Entry Configure Cal Array
:FLATness:POINts
Supported All Models
[:SOURce]:CORRection:FLATness:POINts?
This query returns the number of points in the user- flatness correction file.
:FLATness:PRESet
Supported All Models
CAUTION The current correction data will be overwritten once this command is executed. Save the
current data if needed. Refer to
“:FLATness:STORe” on page 23 for storing user- flatness files.
[:SOURce]:CORRection:FLATness:PRESet
This command presets the user- flatness correction to a factory- defined setting that consists of one
point.
Key Entry Preset List
Range Frequency Standard
Option 1EQ
a
a.Settable, but specified to –127 dBm with Option 1EQ
Option 501
b
: 100kHz–1GHz
b.Option 501 is only available on the N5181A.
–110 to 18 dB <–144 to 18 dB
Option 503: 100kHz–3GHz –110 to 18 dB <–144 to 16 dB
Option 506: 100kHz–6GHz –110 to 18 dB <–144 to 4 dB