56 Agilent N5181A/82A MXG Signal Generators SCPI Command Reference
Basic Function Commands
Power Subsystem ([:SOURce]:POWer)
:STOP
Supported All Models
[:SOURce]:POWer:STOP <value><unit>
[:SOURce]:POWer:STOP?
This command sets the last amplitude point in a step sweep.
*RST 1.35000000E+002
Range Refer to “[:LEVel][:IMMediate][:AMPLitude]” on page 57 for the output power
ranges.
Key Entry Amptd Stop
[:LEVel][:IMMediate]:OFFSet
Supported All Models
[:SOURce]:POWer[:LEVel][:IMMediate]:OFFSet <value><unit>
[:SOURce]:POWer[:LEVel][:IMMediate]:OFFSet?
This command sets the power offset value.
*RST +0.00000000E+000
Range 200 dB to 200 dB
Key Entry Amptd Offset
Remarks This simulates a power level at a test point beyond the RF OUTPUT connector
without changing the actual RF output power. The offset value only affects the
displayed amplitude setting.
You can enter an amplitude offset any time in either normal operation or
amplitude reference mode.