Agilent N5181A/82A MXG Signal Generators SCPI Command Reference 73
System Commands
Display Subsystem (:DISPlay)
:REMote
Supported All Models
:DISPlay:REMote ON|OFF|1|0
:DISPlay:REMote?
This command enables or disables the display updating when the signal generator is remotely
controlled.
ON (1) This choice updates the signal generator display (Text Area) so you can see the
settings as the commands are executed, however, this will degrade the signal
generator speed. Frequency Area, Amplitude Area, and status LEDs continue to
update. For more information on the front panel display description, refer to the
User’s Guide.
OFF (0) This choice turns off the display (Text Area) updating while further optimizing the
signal generator for speed. No Text Area updates occur but the Frequency Area,
Amplitude Area, and status LEDs continue to update. For more information on the
front panel display description, refer to the User’s Guide.
Key Entry Update in Remote Off On
Remarks The setting enabled by this command is not affected by signal generator preset or
*RST. However, cycling the signal generator power will reset it to zero.
[:WINDow][:STATe]
Supported All Models
:DISPlay[:WINDow][:STATe] ON|OFF|1|0
:DISPlay[:WINDow][:STATe]?
This command is used to either blank out (OFF or 0) the display screen or turn it on (ON or 1).
Remarks *RST and presetting the signal generator or cycling the power will turn the display
on.